Usage
10 Maintenance
10.1 Recommended maintenance schedule
Only qualified personnel or personnel trained by Air Techniques is allowed to service the unit�
EN
Prior to working on the device or in case of danger, disconnect it from the mains (e� g� pull the
US
mains plug)�
WARNING
Risk of infection from contaminated products
Risk of cross contamination
Reprocess the product correctly and promptly before its first use and after every subse-
i
quent use�
Inspection in-
terval
Every 3 years
Maintenance
interval
Every 3 years
42
OBJ_DOKU-0000003575-mehrere-005.indd 42
Inspection work
Functional test of the display� Are all symbols displayed?
i
Functional test of the exposure button�
i
Do the various status LEDs light up?
i
Check that the mechanism of the head supports works correctly� Are the
i
head supports easy to detach and put back on?
Functional test of the EMERGENCY OFF button� Is the EMERGENCY OFF
i
button easy to operate mechanically, and does it light up when pressed?
Light barrier test for all light barriers installed in the unit�
i
Visually check the positioning beams� Check the operation of the adjustment
i
lever for the canine positioning beam�
Check the X-ray images for artifacts� If necessary, adjust the collimator and/
i
or calibrate the sensor�
Check the firmware and software versions�
i
Perform a comparative dose measurement based on the requirements from
i
the acceptance test�
Recurring tests and tests after repairs of medical electrical equipment –
i
DIN EN 62353 (VDE 0751-1)�
Maintenance work
Visually and acoustically check the linear movement on the C-shaped arm� If
i
necessary, clean the slide rails with alcohol and grease them with Vaseline�
Check the operation of the lift motor� Does the unit lift and lower with minimal
i
noise? If necessary, clean with alcohol and grease with Vaseline�