Special functions on switching on:
Keys:
Functions:
Setting the desired averaging time
If the "Rep" key is already pressed on switching on the unit
with the "On/Off" key (simultaneously), the indication will
show the maximum averaging time of 7200 seconds.
+
This maximum value can then be reduced in 1800 second steps
with the "Rep" key.
Example: You require an averaging time of 60 min. (3600 sec.)
By pressing the key "Flow" or "Temp" you return to the measuring
mode.
This averaging time remains saved in the "Rep" key until it is
overwritten or the instrument is switched off.
Zero Compensation
Attention: For Zero compensation always mount the
protection cap!
Let the probe acclimate for approx. 15 min. (Temperature
compensation)!
After switching on the unit, press the key "Max" until the decimal
dot flashes. When the dot stops flashing, the zero compensation is
finished.
Attention: For Zero compensation always mount the
protection cap!
Schiltknecht Messtechnik AG
With an indication of 7200 sec., press the key "Rep"
twice (indication 3600)
ThermoAir3
35