INSTRUCTION OF THE USER 
1. 
Identify the biological indicator to be treated. 
2. 
Insert  the  biological  indicator  into  the  sterilizer,  according  to  the 
manufacturer's instructions of the autoclave. 
3. 
Perform the sterilization cycle. 
4. 
Remove the biological indicator from the sterilizer. 
5. 
Let the biological indicator cool for at least 10 minutes. 
6. 
Check the colour change of the chemical strip indicator  (fg.2 – POS. L). 
7. 
Break the glass bulb (fg.2 – POS.M) of the biological indicator which is inside 
the thermoplastic container by exerting pressure with the supplied pliers, until the 
bulb  breaks.  Perform  this  operation  just  a  minute  before  inserting  the  indicator  in 
the  incubator.  The  indicator  must  be  inserted  in  the  incubator  within  and  not 
beyond 1 hour from the moment it was removed from the sterilizer. 
8. 
Insert the biological indicator in the incubator's heating unit  
9.      Begin the cycle of incubation by pressing the START button 
10.    Check  the  biological  indicator  at  regular  intervals  to  control  the  changing  of 
colour (at 18, 24, 36 hours). 
11.  Incubate the biological indicator for at least 48 hours using as a reference an 
indicator not treated in a sterilization cycle.  
12.  For  the  detailed  procedure  carefully  read  the  instructions  for  use  which  are 
inside the packaging of the biological indicators. 
 