Guidance and manufacturer's declaration
Guidance and manufacturer's declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the
user of the Human Care lift should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the
user of the Human Care lift should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
+/- 6 kV contact
discharge (ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power
transient /
supply lines
Burst
+/- 1 kV for input/out-
IEC 61000-4-4
put lines
Surge
+/- 1 kV differential
IEC 61000-4-5
mode
+/- 2 kV common
mode
Voltage dips,
<5 % UT
short interrup-
(>95 % dip in UT)
tions
for 0,5 cycle
and voltage
40 % UT
variations on
(60 % dip in UT)
power supply
for 5 cycles
input lines
70 % UT
IEC 61000-4-11
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip
in UT))
for 5 sec
34
www.humancaregroup.com
Compliance
Group 1
Class B
Class A
Complies
Compliance level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power
supply lines
n/a. for input/output
lines
+/- 1 kV differential
mode
n/a. for common
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
Electro magnetic environment - guidance
The Human Care lift uses RF energy only for its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The Human Care lift is suitable for use in all esta-
blishments including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Electro magnetic environment - guidance
Floors should be wood, concrete or ceramic
tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Human Care lift requires continued
operation during power mains interruptions, it
is recommended that the lift be powered from
an uninterruptible power supply or battery.