Compatibility; Functions; Emergency Lowering - human care HeliQ 55100 Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

HeliQ

Compatibility

Human Care Group slings are designed for use
with all Human Care lifts and hanger bars. The
products are also compatible with products of
other suppliers utilizing the same attachment
method for securing slings to a hanger bar and
lift. However, due to the wide range of products
worldwide, Human Care cannot be responsible
for any errors that may result from improper
use or application of combinations of slings,
hanger bars and lifts of other suppliers. The use
of the combined system is at the prescriber's
own risk and liability.
To ensure the safety of the client and caregiver,
Human Care recommends the following mini-
mum requirements:
1. When using slings with lifts and hang-
er bars, made by manufacturers other
than Human Care, the combined system
needs to be designed with the same sling
attachment method for the hanger bar of
floor lifts and ceiling lifts, meaning loop
style slings are combined with a loop style
hanger bar, clip style slings are combined
with a clip style hanger bar.
2. The involved patient lifts and slings must
be CE marked in accordance with Directive
93/42/EEC or Regulation 2017/745 and
comply with standard requirements for
patient lifts as stated in EN ISO 10535.

Functions

Emergency stop
In an emergency, the emergency stop can be
activated by pulling the red belt. To reset the
emergency stop, push up the activation plate at
the top of the belt.
26
www.humancaregroup.com
3. The caregiver must follow the manufactu-
rer's instruction manuals and recommen-
dations with respect to the use, care,
client, caregiver, cleaning, and inspection
of the slings and lifts.
4. Different products on the assembled lift
system: i.e., lift, hanger bar, sling, scale and
other accessories may have different max.
loads. It is always the lowest maximum
allowable load indicated for the respective
products and the system that applies for
the entire system.
Always check the markings on all individual
products in the system.
5. Individual risk assessment, including a
practical test of the system and validation
with the specific combination used are
required to affirm that the size of the sling
is correct for the intended use and compa-
tible to the width and design of the hanger
bar. The combined system needs to be
documented by the competent assessor.
If there are any questions, please contact your
local Human Care office or distributor.

Emergency lowering

In the event of an emergency, the emergency
lowering button on the lift or hand control can
be pressed to quickly lower the lift. An audio
signal indicates that the lift is lowering. Always
ensure that emergency lowering can be per-
formed as safely as possible for the client. The