10.2 Weekly Maintenance: Cavitron
Select™ SPS™ Ultrasonic Scaler with
Cavitron
Select™ Reservoir System
®
1. Remove Cavitron
Insert from Handpiece.
®
2. Turn the System OFF.
3. Detach Cavitron
®
Select™ SPS™ Scaler from
Reservoir System.
4. Detach Cavitron
Select™ Reservoir Bottle from Pump (see
®
Section 6 for complete instructions).
5. Empty entire contents of Cavitron
6. Replace empty Cavitron
7. Carefully lock Cavitron
Cavitron
®
Select™ Reservoir Pump.
8. Connect Cavitron
®
Select™ SPS™ Scaler to Cavitron
Select™ Reservoir System (see Section 6 for complete
instructions).
9. Turn the System ON.
10. Hold the Handpiece (without the Cavitron
upright over a sink or drain.
11. To air purge the water line, operate the Cavitron
SPS™ Scaler for at least 30 seconds. (This will eliminate
fluid to prevent it from settling in the Cavitron
SPS™ Scaler for extended periods of time. This may also
help decrease the chance of bio-film formations in the
Cavitron
Select™ SPS™ water lines.)
®
12. Turn the System OFF.
13. Prepare a 1:10 hypochlorite (NaOCL) solution by mixing
1 part 5.25% sodium hypochlorite with 10 parts tap water.
14. Remove cap and pour this 1:10 solution into the Cavitron
Select™ SPS™ Reservoir Bottle. Do not overfill with solu-
tion (see Section 3.3 for complete instructions).
15. Securely replace cap on Bottle.
16. Turn the System ON.
17. Hold the Handpiece (without the Cavitron
upright over a sink or drain.
18. Operate the Cavitron
®
10 seconds.
19. Turn the System OFF.
20. Leave the unit undisturbed for at least 10 minutes, BUT
FOR NO MORE THAN 30 MINUTES to allow sodium
hypochlorite solution to disinfect the system. As a suggestion,
place a sign which states that the System is being disinfected
place a sign which states that the System is being disinfected
and should not be used.
21. Unscrew cap and discard disinfectant solution.
22. Fill Cavitron
®
Select™ Reservoir Bottle with tap or distilled
water. Do not overfill with liquid (see Section 3.3 for
complete instructions).
23. Turn the System ON.
24. Operate the Cavitron
®
60 seconds.
25. Turn off system. Detach Cavitron
from Select™ Reservoir System.
®
®
Select™ Reservoir Bottle.
Select™ Reservoir Bottle to Pump.
®
Select™ Reservoir Bottle into
®
Insert installed)
®
®
Select™
®
Select™
®
Insert installed)
Select™ SPS™ Scaler for at least
Select™ SPS™ Scaler for at least
®
Select™ SPS™ scaler
26. Detach Cavitron
(see Section 6 for complete instructions).
27. Autoclave Cavitron
NOTE: ALL CHEMICALS MUST BE FLUSHED FROM THE
SYSTEM BEFORE IT IS READY FOR PATIENT USE.

10.3 Water Line Filter Maintenance

When the water line filter becomes discolored, the filter should be
replaced to prevent reduced water flow to the Cavitron
SPS™ System. A 10-pack of replacement filters is available
by ordering Part Number 90158 from your local DENTSPLY
by ordering Part Number 90158 from your local DENTSPLY
by ordering Part Number 90158 from your local DENTSPLY
distributor.
1.
Disconnect the water supply hose from the water source. If
a quick-disconnect connector is attached to the end of the
®
hose, relieve the water pressure by pressing the tip of the
connector in an appropriate container and drain the water.
2.
Grasp the fittings on either side of the filter disk and twist
counterclockwise. Remove the filter section from either side
of the water hose.
3.
Install the replacement filter onto the water hose fittings.
The filter should be positioned to match up with the correct
hose fitting.
4.
Hand tighten the two hose fittings in a clockwise direction.
Reconnect the water supply hose, operate the unit to bleed
the air and test for leaks.
10.4 Pump Lavage Filter Maintenan ce
When the pump lavage filter becomes discolored, the filter should
®
be replaced to prevent reduced water flow to the Cavitron
Select™ SPS™ System. A 10-pack of replacement filters is
available by ordering Part Number 90191 from your local
DENTSPLY
DENTSPLY
DENTSPLY distributor.
®
®
1.
Empty the contents of the Cavitron
bottle and disconnect it from the pump.
2.
Remove the old filter and discard.
3.
Insert the small end of the new filter until filter is seated.
10.5 O-Ring Maintenance
Before placing Cavitron
the following steps will increase the O-ring life on Cavitron
Inserts:
• Fill entire handpiece with water.
• Lubricate O-ring (black or green rubber material) with water.
• Gently twist the insert down into the handpiece until the insert
is fully seated in place.
Section 10: System Care cont'd

System Care

System Care cont'd
Select™ Reservoir Bottle from Pump
®
®
Select™ Reservoir Bottle.
Select™ Reservoir
®
Inserts into the handpiece, performing
®
ENGLISH • 13
®
Select™
®
®
®
®