Attach the interface cable
§ Pin Assignments: see "Pin Assignment Chart" section.
§ Attach the connecting cable on the peripheral device to the indicator with the cable
gland.
§ Prepare the interface cable:
– Strip the isolation on the cable end by approx. 10 cm
– Shorten the shield to approx. 1 cm and pull back over the insulation
– Strip approximately 1 cm of the insulation from the connecting cable wires and affix
ferrules to the wire ends.
§ Attach the cable gland:
!Please use extreme caution when performing any work on the equipment that affects this
cable gland. Use a torque wrench. Torque of this cable gland: 5 Nm.
– Remove the protective cap from the bore hole on the indicator
– Push the supplied cable gland through the bore hole and secure from the inside using the
locknut (1).
– Push the cable through the cable gland until the shield (2) makes contact with the clamps
(3). Tighten the compression nut (4) until the sealing insert (5) between compression nut
and cable forms a small bead.
– Check the contact between the shield and the clamps.
§ Attach the connecting cable:
– Securely connect the wires in accordance with the terminal assignment charts.
– For assignment charts see page 16.
§ Close the indicator again. Make sure that the sealing ring between the housing and the
front panel is correctly seated.
§ After completing the installation works, check the IP69K protection. Use a pressure gauge
for this purpose. Contact the Minebea Intec Service Center for further details.
Setting the UniCOM interface for operation
After installing the interface module in the indicator and configuring it, set the UniCOM
interface or "controller" for the intended function with the aid of the operating menu.
See the "UniCOM Settings" section for this purpose. The "Configuration" section of the
operating instructions contains further explanations about this.