1. Intended Use / Description of the Function

®
1.1 The AEROsuc
person without the need for a power supply.
1.2 It is used in medical technology, especially in emergency medicine, for
extracting accumulated blood, mucus, saliva, etc. as well as viscous food
components.
1.3 This eliminates obstruction of the upper airway and the risk of aspiration.
®
1.4 The AEROsuc

2. Operation / Handling

2.1 Select a suction tube according to the size of the patient (size L - adult | size
S - child).
2.2 Fit the tube or the adapter to the suction receptacle.
2.3 Fix the receptacle to the suction pump.
Caution: It is particularly important to ensure that the receptacle is fitted
properly to the housing of the pump to make the system air tight and allow a
vacuum to be created.
®
2.4 The AEROsuc
handle of the pump.
2.5 A low pressure condition is produced by the pumping movements and the liquid
is sucked out of the patient's mouth and throat through the suction tube into the
secretion receptacle.
2.6 The lever returns automatically each time to the starting position.
2.7 Blood, secretions and mucus as well as viscous food components can be
extracted directly via the suction tube (size S | size L) or by directly connecting a
suction catheter by means of the supplied adapter for funnel connectors.
2.8 The suction receptacle has a capacity of 300 ml. It is equipped with an overflow
protection mechanism (protects the pump from contamination) and a valve
(prevents reverse flow to the patient). These mechanisms prevent liquids
and particles from getting into the vacuum pump or returning to the patient.
Caution: If the pump is turned over, overflow protection is activated immediately
making suction impossible. The pump handle and the lower surface of the
Rescue is a manual suction pump that can be used by a trained
Rescue is not intended for bronchial suctioning.
Rescue is operated by pulling the lever repeatedly towards the

English