7. Change of the drill
Switch off mains switch. Press operation plate (7) upwards. Remove old drill
- use finger guard!-
Insert new drill.
Max. tool lenght 34.0 mm! Release operation plate into zero position. Adjust the drilling
depth gauge. Switch on the unit.
Unit is now prepared for processing.
OFF
7
8. Maintenance and care
Clean device occasionally with a soft cloth.
When using without suction open Cover of the drilling table and vacuum the inside aerea
occasinally. Adjusting of the laser pointer by opening the hexagon screw with hexagon key
SW03. After positioning close screw again (Pic. 2 - 4)
Further maintenance is not necessary!
Never use compressed air or cleaning agent!
(Abb. / Pic. 1)
9. Possible faults
Net switch lighting and net LED are not working:
- > net fuse( 3,15 A/T 220V No.1396)
(6,3A/T 100V-115V No.3107)
- check and exchange when required!
By all other faults send the device directly to SCHICK-Dental or an authorized dealer
(Abb. / Pic. 2)
!
(Abb. / Pic. 3)
(Abb. / Pic. 4)