Télécharger Imprimer la page

EKOM DO 2.1 Manuel De L'utilisateur page 12

Publicité

DO 2.1
DO 2.1-10
12. WIRING DIAGRAM
230 V
ELECTRIC OBJECT OF
CLASS I
Mark
Name
M1
Suction pump
EV1
Fan
CP1
Switching block
13. FIRST OPERATION
 Make sure that all stabilizers used during transport were removed.
 Check all connections of the vacuum and exhaust pipe-work system.
 Check the suction unit is switched on at the plug.
 Check the suction unit switches on at the dentist's pneumatic control module.
OPERATION
In case of emergency, disconnect the suction unit from the mains (pull out the mains
plug).
Parts of the suction unit can have hot surface areas. There is the danger of scorching if
contact is made.
The suction unit operates after receiving a signal from the dental chair. The cooling fan
and suction pump always work together.
14. SWITCHING SUCTION UNIT ON
When the suction unit is switched on at the plug, power is provided to the switching block, and puts the
aspirator - on stand-by mode - waiting for a signal from the dental chair. When a signal is received, the
suction unit operates and creates a vacuum in the inlet path until a signal interrupts its operation, and puts
the suction unit back - on stand-by mode.
08/2018
PST-DO2
X2
5
4
Re1
Tr1
FU1 FU2
+
C1
D1
X1
1
2
U
N
PS-DO2
X2
5
4
Re1
+
FU1 FU2
C1
D1
X1
1
2
U
N
- 9 -
EV1
M1
C
M
B1
1
CP1
X2
5
4
PS-DO2
(PST-DO2)
X1
1
2
U
N
24V AC/DC
U
N
PE
or EXT.
NP-DO2.1_10_08-2018