STEP 49
CH2 DC offset at + full scale.
CH2 DC offset at full scale.
STEP 50
STEP 52
CH2 Multiplier offset.
STEP 51
CH2 Multiplier zero.
STEP 53
CH2 Waveform offset Filter 1.
STEP 54
CH2 Output level at fullscale Filter 1
STEP 55
CH2 Waveform offset Filter 2
STEP 56
CH2 Output level at fullscale Filter 2
STEP 57
CH2 20dB attenuator (A)
STEP 58
CH2 20dB attenuator (B)
STEP 59
CH2 10dB attenuator
STEP 60
CH2 Sine amplitude flatness
STEP 61
CH2 Level 1MHz
STEP 62
CH2 Level 2MHz
STEP 63
CH2 Level 5MHz
STEP 64
CH2 Level 10MHz
STEP 65
CH2 Level 15MHz
STEP 66
CH2 Level 20MHz
STEP 67
CH2 Level 25MHz
STEP 68
CH2 Level 30MHz
STEP 69
CH2 Level 35MHz
STEP 70
CH2 Level 40MHz
STEP 71
CH2 Level 43MHz
STEP 72
CH2 Level 45MHz
STEP 73
CH2 Level 47MHz
STEP 74
CH2 Level 50MHz
STEP 75
CH2 Square-wave high level
STEP 76
CH2 Square-wave low level
STEP 77
CH2 Rise time (0)
STEP 78
CH2 Rise time (0)
STEP 79
CH2 Rise time (1)
STEP 80
CH2 Rise time (1)
STEP 81
CH2 Rise time (2)
STEP 82
CH2 Rise time (2)
STEP 83
CH2 Fall time (0)
STEP 84
CH2 Fall time (0)
STEP 85
CH2 Fall time (1)
STEP 86
CH2 Fall time (1)
STEP 87
CH2 Fall time (2)
Adjust for + 10V ± 10mV
Check for –10V ± 50mV
Adjust for 0V ± 3mV
Adjust for minimum Volts AC
Adjust for 0V ± 3mV
Adjust for 10V ± 10mV
Adjust for 0V ± 3mV
Adjust for 10V ± 10mV
Adjust for 1V ± 1mV
Adjust for 1V ± 1mV
Adjust for 3.162V ± 3mV
Note reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for 10V ± 10mV
Adjust for -10V ± 10mV
Adjust for 50ns
Adjust for 100ns
Adjust for 100ns
Adjust for 2us
Adjust for 2us
Adjust for 40us
Adjust for 50ns
Adjust for 100ns
Adjust for 100ns
Adjust for 2us
Adjust for 2us
93