WARNINGS ON INTERFERENCE RISKS DURING USE IN DIAGNOSTIC
INVESTIGATIONS
This device is designed to meet the current requirements for electromagnetic compatibility.
As far as EMC requirements are concerned, electromedical devices require special care
when being installed and used, and are therefore required to be installed and/or used
in accordance with the manufacturer's specifications. Risk of potential electromagnetic
interference with other devices. Mobile or portable RF radio and telecommunication
devices (mobile phones or wireless connections) may interfere with the operation of
electromedical devices. The device may be susceptible to electromagnetic interference in
the presence of other devices used for specific diagnosis or treatment. For more information
visit www.flaemnuova.it.
TROUBLESHOOTING
Problem
Cause
• Malfunction of the
medicine tray and
mesh unit.
• The medicine does
not flow into the
The device switches on but
nebulising area.
does not nebulise or does
it poorly.
• The nebulising
speed depends
on the type of
medicine used.
• Liquid has seeped
out from the mesh
• No liquid in the
The blue LEDs flash and the
tray.
appliance switches off after
• Five minutes have
few seconds.
elapsed from
The appliance switches off.
switch-on.
Light button (1A) turns red
and flashes and the appli-
• Batteries are down
ance does not nebulise or
or depleted.
switches off.
Remedy
• Immerse the medicine tray and
mesh unit (2) in a solution of water
(50%) and white wine vinegar, rinse
and reassemble as described in
section 6.3 SANITISATION.
• Ensure the small hole in the centre
of the medicine tray cover (2B) is
not obstructed, if so, clean it.
• The inhalation treatment time de-
pends on the type of medicine and
on the patient's inhaling capacity.
• Angle: it must be kept upright when
in use. If it is tilted backward, the
medicine does not touch the mesh.
• Gently dry the mesh with a cotton
swab or shake to remove any drop-
lets from the medicine outlet area.
• Check for liquid in the medicine tray
(2).
• Switch the appliance on again.
• Replace the batteries (6) if alkaline.
• Recharge the Batteries if recharge-
able Ni-MH or Lithium.
23