Technical data according to DIN 12876
max. delivery pressure
Delivery at 0,5 bar
Delivery at 1,0 bar
Delivery at 1,5 bar
Delivery at 2,0 bar
Delivery at 2,5 bar
Pump connection
max. permissible kin. viscosity
Cooling water connection
Consumption at water 15°C, flow 0°C
min. cooling water differential pressure
max. cooling water pressure
min. filling capacity
Filling capacity expansion tank
Overall dimensions WxDxH **
Net weight
Power supply factory configured (3 Phase)
max. current (3 Phase)
Fuse (3 phase)
Pressure equipment category
Degree of Protection
min. ambient temperature
max. ambient temperature
from Serial-No.:
Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Included Accessories:
mini-USB cable #54949, E-grade "Professional" #9496, hose connection for G1/2 male,
Optional accessories:
E-grade "Explore" #10495, SpyLight-Software, Com.G@te Namur, PC-Com.G@te-cable, Holder for Com.G@te #10018, Com.G@te-
extension cable: upon request, RS232 adapter cable #55018, Thermofluid, external pressure sensor, metal hoses, braided hoses for
cooling water, external sensor, connecting cable, isolation sleeve for external open applications, float switch in sight glass for extended
security, further accessories, etc.: see catalog.
Note: Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN
3863, sleeve nut according to DIN 3870.
Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -
outlet. This temperature control unit has been designed to operate with cooling water up to 20°C. As the cooling water temperature
increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible. Materiels used in the
cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.
in accordance with EN60034-1 the following voltage and frequency tolerances are valid:
Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%
Example
-5% voltage and +2% frequency -> not allowed!
-5% voltage and - 2% frequency -> allowed
Information to Electromagnetic compatibility:
Classification (disturbance) to EN55011: Class A, Group 1
Attention: leakage current > 3,5mA
Standard delivery conditions - Power cable configuration:
1. Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)
2. Three-phase devices with current consumption less than 63A --> with cable, without plug
3. Three-phase devices with current consumption greater than 63A --> without cable, without plug
This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process
refrigeration. Certification by a Notified Body upon request.
** Please respect space requirements. See operating conditions at www.huber-online.com
Peter Huber Kältemaschinenbau SE
3 bar
61 l/min
55 l/min
48 l/min
39 l/min
28 l/min
M30x1,5 male
50 mm²/s
G1/2 male
130 l/h
1 bar
6 bar
3,2 l
3,4 l
460x604x1465 mm
231 kg
400V 3~ 50Hz
12,5 A
3x16 A
I
IP20
5 °C
40 °C
535218
Werner-von-Siemens-Str. 1
D-77656 Offenburg
Tel 0781/9603-0
Fax 0781/57211
1.1/24