Refrigerated Heating Circulator with water-cooled cooling machine and optical level indicator. Magnetic coupled circulation
pump made of stainless steel. Automatical switch-over and capacity adaption for heating and cooling machine. Copper
soldered evaporator, moistened parts and housing made of stainless steel. As well as for externally closed and also
externally open applications. With adjustable overtemperature protection according to DIN 12876. Powerful variable speed
pump (soft start) with integrated pressure control with optional external pressure sensor.
Unistat „P" Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the
Flow-Through chemistry or in the Semicon industry.
Pilot ONE:
The new Pilot ONE controller with pioneering technology and advanced control functions brings numerous advantages to
routine work. The extensive features list includes a brilliant 5,7" TFT touchscreen display, USB and network connections,
an integrated technical glossary and language support in 13 languages (EN, DE, FR, IT, ES, RU, CN, PT, JP, CZ, PL, KO,
TR). The Pilot ONE has a convenient navigation system with easily remembered icons and menu categories which are colour
sorted to make routine work simpler. Thanks to a favourites menu and One-Click operator guidance all important information
is always just a few keystrokes away. Software wizards also help you to set up, ensuring correct settings. The USB port
allows connection of the system to a PC or notebook. Together with the Spy software, requirements such as remote
control or data transmission are easily achieved in a cost-effective manner. Network integration is easy with the internet
port.
further functions:
E-grade Professional installed as standard, TAC (True Adaptive Control) - self optimising internal and cascade control,
selectable temperature control mode (Internal/Process), programmer with 10 programs (max. 100 steps), ramp function
(linear and non-linear), 5 point calibration, scalable graphic display, favourites menu, display resolution 0,01 K, integrated
technical glossary, 2nd set point, user menus (Administrator level), calendar start, wallpaper selection.
4-year warranty - registration required.
Technical data according to DIN 12876
Operating temperature range
Temperature stability at -10°C
temperature set point / display
Resolution of display
Internal temperature sensor
Sensor external connection
Interface digital
digital input
digital output
Alarm message
Safety classification
Heating power
Cooling power with
at 250°C
at 200°C
at 100°C
Cooling power with
at 0°C
at -20°C
at -40°C
at -60°C
at -80°C
Refrigeration machine
Refrigerant (ASHRAE, GHS)
Global Warming Potential (GWP)
Refrigerant quantity
Refrigerant 2nd stage (ASHRAE, GHS)
Global Warming Potential (GWP)
Refrigerant quantity 2nd stage
Circulation pump:
max. delivery
Peter Huber Kältemaschinenbau SE
Unistat P815w
-85...250 °C
0,01 K
5,7" colour Touchscreen
0,01 K
Pt100
Pt100
Ethernet, USB (Host u.
Device), RS232
ECS ONE
POKO ONE
optic, acoustic, relay
III / FL
2 kW
Thermooil
1,5 kW
1,5 kW
1,5 kW
Ethanol
1,5 kW
1,5 kW
1,4 kW
1,2 kW
0,2 kW
water-cooled, CFC- and
HCFC-free
R-452A (A1, H280)
2141
0,95 kg
R-23 (A1, H280)
14800
0,39 kg
MK pump
67 l/min
Werner-von-Siemens-Str. 1
D-77656 Offenburg
Order-No.:
1053.0010.01
Tel 0781/9603-0
Fax 0781/57211
www.huber-online.com