MAINTENANCE AND CLEANING
Warning! Every three months, check the
product for wear and deflated cells. In case of
serious damage, or a large number of deflated
cells, do not continue to use the product.
Consult your Assistive Technology Professional.
Warning! Make sure the SmartCell liner is
placed in the correct compartment.
Caution! Do not dry clean, steam clean or
autoclave the product. The product cannot
withstand excessive heat or pressure. When the
given cleaning instructions are not followed, it
is very likely that the product and/or the
product's cover will age more rapidly.
Caution! Do not use aggressive or abrasive
cleaners, or sharp objects to clean the cushion.
Caution! Remove foam inlay before washing
this cushion in a washing machine.
Caution! Do not use the foam inlay without the
foam inlay liner.
Caution! Do not place the topside of the Vicair
O2 cushion against the glass of the washer
door.
Caution! Excessive washing may influence the
aging of this product.
Cleaning the Vicair Active O2
After you have emptied the storage pouch and
removed the foam inlay, your Vicair Active O2
cushion can be machine washed at 60°C / 140
°F using an anti-crease cycle and mild
detergent. You do not need to remove the
SmartCell liner and SmartCells before putting
the cushion into the washing machine. Make
sure the topside of the cushion is not placed
against the glass of the washer door. Do not
load the machine with more than half of its
maximum load weight. Spin speed should be
reduced. When using a machine with agitator,
make sure that the cushion can move freely
within the drum. Do not use any chlorine bleach
and do not dry clean, steam clean or autoclave!
Do not tumble dry your Vicair Active O2 cushion
but let it air dry (5-6 hours).
Cleaning the Vicair Active O2 foam inlay
Do not machine wash the foam inlay and do not
immerse it in water or any other liquid. To clean
the foam inlay you first remove it from its
compartment.
Cleaning the foam inlay liner
Take the foam out of the foam inlay liner.
The emptied liner can be machine washed
at 60°C. Follow the washing instructions on
the label attached to the inside of the liner.
Cleaning the foam
Take the foam out of the foam inlay liner
and clean it by using a disinfectant spray.
Disinfecting your Vicair Active O2
Warning! In case of soiling with body fluids, and
when transferring the product to another user,
the product must be disinfected to avoid cross
contamination. Clean the Active O2 cover, foam
inlay and SmartCell liner as instructed above.
Preferably use an anti-bacterial detergent. After
cleaning spray your cover, foam inlay liner and
SmartCell liner with a disinfectant. Bend the
cover compartments outwards to make sure the
spray gets into the gaps.
(Re)assembling your Vicair Active O2
The Vicair Active O2 is fully assembled on
delivery. When it needs to be (re)assembled, for
7