Alarm Battery
Maintenance
Unit comes with (4) AA Alkaline 1.5 V batteries. The back-up
time for the alarm battery system is 60 hours. Battery must be
replaced after an active alarm or after every 12 months,
whichever is the earliest.
Steps to follow for battery replacement:
1. Locate the knurled/thumb screw on the left side of the
header panel.
2. Remove the knurled/thumb screw by hand.
3. Once the screws are removed, wear hand gloves and
gently pull the bracket out until you see the batteries
completely.
4. Replace the batteries and gently close the bracket. Once
the bracket is closed, align the holes on the bracket and
header panel, screw using the knurled/thumb screw.
20 |
Maintenance
Preparation for Storage
If the unit is going to be stored in an off condition, allow the
unit to warm up and dry out with the door open before moving
into storage.
When the freezer is placed in intermediate storage, make sure
that the ambient temperature is between 15°C and 32°C
(59°F to 90°F) and the maximum relative humidity does not
exceed 60 %, non-condensing.
General Purpose Freezers