EN - 12
4. Operation
4.1. Start a Simple Measurement with MB92
1. Switch on the instrument by pressing
2. Configure testing parameters
a. Tap on
b. To edit an existed method, tap on a method >Edit.
Or you can create a new method by tapping on New > Name the method
> Save
c. Select a Drying Program.
d. Set drying temperature. The temperature range is 40 -200 ℃.
e. Set Switch-off criterion
◦ Auto: Shut off drying if the weight fluctuation is less than 1mg in required
seconds, e.g. A60 means "in 60 seconds".
◦ Manual: Shut off drying manually by pressing
◦ Timed: Shut off drying after the preset drying time is elapsed.
Note:
After a new method is created, the method will not be selected for
testing automatically. To select this new method, tapping on
the new Method > Load.
3. Place sample pan and close the lid to tare. Tare is performed automatically
when the lid is closed.
4. Open the lid and add sample.
5. Close the lid to start drying.
Note:
Drying will start automatically when the lid is closed. To disable auto-
drying, navigate to
Manual.
icon to enter the Method menu
> Your Method > Edit > Starting mode >
MB92 Quick Start Guide
.
.
>