If the AB myPilot stops being transmitting commands to the Naída CI, it might be
necessary to re-pair the AB myPilot with the Naída CI. Follow the steps in the Pairing
the AB myPilot section of this user guide.
•
If volume commands from the AB myPilot to the Naída CI seem erratic, position the
AB myPilot within 30 cm (12" ) of the Naída CI. If the problem persists, re-pair the
AB myPilot and the Naída CI.
•
The digitally-coded, inductive transmission technology used in the AB myPilot
prevents interference from most other devices. When operating the AB myPilot near
a computer terminal or other strong electromagnetic fields, it may be necessary to
be at least 60 cm (24") away to ensure proper operation. If the Naída CI does not
respond to the AB myPilot because of an unusual field disturbance, move away
from the disturbing field.
•
Do not use the AB myPilot in locations where it is forbidden to use
electronic devices.
•
X-ray radiation (e.g. CT scans, MRI scans) may adversely affect the correct
functioning of the AB myPilot. We recommend that you turn the AB myPilot off
before undergoing X-ray procedures and leave it outside the room.
4