Situation
AB myPilot
commands apply only
to one Naída CI as
opposed to both
AB myPilot
commands are
not received by
the Naída CI(s)
Most Probable Cause
Naída CIs may be turned
OFF
Naída CI PowerCel
may be depleted
AB myPilot may be
out of range of Naída CI
AB myPilot is too close
or too far away from
Naída CI(s)
Remedy
Naída CIs must be turned
ON to correspond with AB
myPilot. Turn them ON by
attaching PowerCel
to processor
Charge PowerCel or
replace Zn-Air batteries
Move AB myPilot closer
to unresponsive Naída CI
Hold AB myPilot > 1cm
to 30cm (1/2" -12") from
Naída CI and try again
27