Battery status, charger station
The battery symbol lights green when battery capacity exceeds 50 %.
If the battery symbol blinks orange when the lift unit is started, the battery needs to be charged.
If the battery symbol lights orange and a beep is heard when the lift unit is run up, the battery needs to
be charged immediately.
When the lift unit is started the battery symbol lights up for 2 seconds to indicate the current battery
status.
Six seconds after every operation the battery symbol lights for 2 seconds to indicate the current
battery status. Only hand control charging or end point charging mode.
Charging indication (hand control charging and end point charging)
When the hand control is placed in the charger station 2 beep sounds are heard and
the green light should stay on for 15 minutes.
Battery status, in-rail charging
The battery symbol lights green when battery capacity exceeds 50%.
When the lift unit is activated the battery symbol lights green for 15 minutes.
If the battery symbol and the book symbol blink orange when the lift unit is started, the circuit for the charging
system has been broken and must be checked by a technician.
Active charging detection (In-rail charging)
After activation the lift unit will light the battery symbol for 15 minutes and after 4 minutes the charger should turn
on. If the lift unit is missing the charger it will start indicate after 4 minutes. Green 15 minutes after activation.
NOTE!
Use only the charger that is intended for the lift unit.
For maximum service life, charge batteries regularly. Do not allow the battery charge level to fall below 25%
(orange symbol on control panel).
Emergency stop must not be activated during charging.
8
Batteries may only be replaced by authorized personnel.