Technical Specifications
Name of product
Model
Frequency
Output current
Weight
Dimensions
Power consumption
AC adaptor
Power source
Input (adaptor used)
Output
Service Life of Device
Period of time specified by the manufacturer during
which the device is expected to remain safe for use.
Applied parts
Part of Revitive that in normal use come into physical
contact with user to perform therapy
24
5827_IFU01_17102527.indd 24-25
5827_IFU01_17102527.indd 24-25
Revitive Essential
RLV
20Hz – 53Hz
Max 8.5mA
1.13kg
Ø 355 mm x 77(D) mm
5W
CE Approved
100-240V
AC ( ), 50/60Hz, 0.18A
5V (
) DC
,1.0A
4 Years
Foot-pads - 135.5cm
2
Electrode Body Pads - 5cm x 5cm = 25cm
2
Output Specifications for Electrical Muscle Stimulation (EMS):
Waveform
Shape
Maximum Output Voltage:
@500Ω
Pulse Duration:
Net Charge @ 500Ω
Maximum Power Density @ 500Ω
ON Time
OFF Time
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including
any d.c. component, shall not deviate by more than ± 20 % when measured with a load
resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance after a
certain period of time, contact the manufacturer
EN
Monophasic
Square symmetrical with polarity reversal
Foot: 36V
Body: 19V
Foot: 480-505µs or
Body: 470-500µs
1070µs
[0.001]mC
Foot: 0.68 mW/cm²
Body: 1.66 mW/cm²
Foot: 1.93 - 8.37s
Body: 1.93 - 6.53s
Foot: 1.00 - 1.50s
Body: 1.00s