To avoid
operation.
Check the chemical resistance to the liquid and to any possible cleaning agents.
The flow meter must be installed without any mechanical tension.
When installed stationary within a plant, fit the flow meter into the pipework with pipe
connections on both sides. In the event of a defect, the flow meter can be simply and
quickly removed and re-fitted.
A T-section should also be used downstream of the second pipe fitting on the flow
meter to
1. Allow for calibration without having to dismantle it.
2. Shut off the pipe work so that the unit can be removed with ease.
Pressure surges that are greater than the nominal pressure stated
on the specification plate (see technical data) can damage the flow
meter.
Important:
Pressure surges are produced by mass forces (content of long
pipes) that are caused by quick-closing fittings!
Keep organic solvents away from the keyboard and LCD display.
24 / 60
contamination, we
recommend a cleaning prior to initial
Operating instructions FMC