Storage and transportation
• The temperature for transportation and storage of the unit must be
between 10 and 50 degrees Celsius.
• The unit must always be transported in its original packaging.
• Avoid subjecting the packaging to shocks and impacts.
• Do not use the product if the packaging is broken or damaged.
Troubleshooting
Symptom
No power .
The shower trolley is
skewed even though
the tilt function is not
activated .
The shower trolley
cannot be raised or
lowered .
Contact Pressalit A/S or your local dealer if the problem cannot be remedied. Do
not attempt to repair the product yourself, as this will invalidate the guarantee.
Expected service life of the shower change trolley is 10 years.
Cause
Battery depleted.
The lifting
columns are out
of sync.
If a statically
charged person
discharges near
the shower
trolley control
box, a fault may
occur in the
lifting columns.
Action
Recharge the battery.
See "Charging the battery".
Reset the lifting columns. See
"Resetting".
Reset the lifting columns. See
"Resetting".