8. The results will be displayed.
9. Choose an option.
¬ Tap
to add a tag.
¬ Tap the highlighted
¬ Tap the hightlighted
¬ Tap the highlighted
¬ Tap
to return to the Test Screen.
10. REMOVE THE DISK FROM THE CHAMBER. DISPOSE OF THE DISK. Disks left in the chamber
can leak and damage the device. Do not store the disk cover on the hub.
11. Press and hold
For the most accurate results samples should be 50 – 104 °F (10 – 40 °C).
The blank well allows the readings to be corrected for small amounts of color and turbidity in the
sample water. For the best results allow samples with a large amount of solids to settle before
testing.
NOTE: For water samples over 100 °F (38 °C) subtract 0.1 from pH result or, for the most
accurate result, wait until water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results. See Cleaning.
SALTWATER SAMPLE DILUTION Test results that are out of range of a reagent system will be RED.
If results for Nitrate, Nitrite, Ammonia or Phosphate are out of range, the water sample must be
diluted with saltwater to obtain accurate results. The salinity concentration of the sample must
be maintained for the reagent system to function properly. The water sample cannot be diluted
with tap water or water that does not contain salt.
Preparation of the Saltwater
1. Prepare 1 liter of synthetic seawater in distilled or deionized water according to the
manufacturer's instructions.
2. Test the prepared seawater with a saltwater Spin disk. Compare the results to the
manufacturer's specifications for Alkalinity, Calcium, Magnesium, and pH. The results
should meet the manufacturer's specifications. The results for Nitrate, Nitrite, Ammonia and
Phosphate should be 0 ppm.
Dilution Procedure If a test result for Nitrate, Nitrite, Ammonia or Phosphate is RED, the result
is out of the range of the reagent system and must be diluted and retested to obtain a reading
which is in the concentration range for the test.
The sample water and saltwater can be measured with any accurate measuring device
(graduated cylinder, pipet, or measuring spoon) if the ratio of sample water to salt water is
maintained. For example, mix 1 mL of sample water and 9 mL of saltwater or mix 1 teaspoon of
sample water and 9 teaspoons of saltwater.
1. Add 1 mL of sample water to a container.
2. Add 9 mL of saltwater to the container.
3. Mix by swirling or stirring with a clean stirrer.
4. Test the diluted sample.
5. Multiply the results on the screen for Nitrite, Nitrate, Ammonia, and Phosphate by 10.
Results for Calcium, Magnesium and pH are not valid for the
diluted sample.
¬ For a smaller dilution, mix 1 mL of sample water with 1 mL
of saltwater. Multiply the test result by 2.
¬ For a larger dilution, mix 1 mL of sample water with 19 mL
of saltwater. Multiply the test result by 20.
¬ Ranges Enabled must be ON for the out of range test
results to be displayed in RED.
SAVE, PRINT, AND SEND TEST RESULTS
Test results can be saved, transferred to the WaterLink Connect
2 mobile app via Bluetooth, and sent to the BLE Mobile Printer
(5-0067).
Test History Settings
The WaterLink Spin Touch can log test results for 250 water
to save the test results to the test log if Auto Save is not enabled.
to send the results to the enabled Mobile Bluetooth Printer.
to send the results to a Bluetooth enabled device.
for 2 seconds to turn the meter off.
Tagging
5
Test Results
Test Time 07:11:34 16-Apr-2020
ALK 133
AMMO 0.5
G HARD 77
NITRITE 1.1
pH 5.7
NITRATE 23
PHOS 0.6
Save
Print
Transfer
Results
Results
Results
via
Bluetooth
(Not Available)
Return
to Test
Screen