Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Enclosure Door
Phenomenon
ELECTROSTATIC
DISCHARGE
Radiated radio-frequency
electromagnetic fields
Fields near radio-
frequency devices for
wireless communication
Electromagnetic fields for
RATED mains frequency.
) The device is intended for use in a home environment or professional healthcare structure and may
a
exclusively be used in hospital rooms and in sections of hospitals or clinics set up for respiratory therapy. The
strictest limits for acceptance were considered and applied during immunity testing.
b
) Before modulation.
) This test level presupposes a minimum distance of at least 15 cm between the AirProjet plus
c
02038204100000 and the sources of magnetic fields at the mains frequency.
Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Door for access with
Phenomenon
Fast electrical transient/
burst
Line to line voltage
impulses
Line to ground voltage
impulses
Disturbances conducted
induced by radio-
frequency fields
Applicable EMC policy
or test method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
alternating current
Applicable EMC policy
or test method
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
18
Immunity test levels
Professional
healthcare
environments
± 8 kV in contact
± 2 kV, ±4 kV ±, ±8 kV, ±15 kV air
)
a
COMPLIANT
NB: further information about the distances to be
maintained from portable and mobile radio-frequen-
cy communication devices (transmitters) and AirPro-
jet plus
02038204100000 may be requested by
contacting PIKDARE S.r.l. using the contact informa-
tion included in this booklet. It is nevertheless advis-
able to keep the electromechanical aerosol therapy
device at a safe distance (at least 0.5 m) from mobile
phones or other communication devices that use ra-
dio frequencies, to minimise interference.
30 A/m
c
)
50 Hz or 60 Hz
Immunity test levels
Professional
healthcare
environments
± 2 kV
Repetition frequency at 100 kHz
± 0.5 kV, ±1 kV
Class II Device or equipped with Class II AC/DC
adapter. Test not applicable.
± 0.5 kV, ±1kV; ±2kV
3 V
) 0.15 MHz - 80 MHz
a
6 V
a
) for ISM bands ranging from 0.15 MHz to
80 MHz
)
b
80% AM at 1 kHz
Home environment
10 V/m
b
)
80 MHz - 2.7 GHz
80% AM at 1 kHz
Home environment