Use of the mask 2 in 1 for adults and children:
• Connect the mask (A) to the mouthpiece (B); then connect the mouthpiece to the corrugated tube (D);
Use of the nose fork:
• Connect the nose fork (C) to the corrugated tube (D);
• Lift the nozzle cap (E);
• Connect the tube to the nebulisation nozzle (G);
• Connect the detachable power cord to the unit;
• Connect the plug of the detachable power cord to the mains;
• Start the unit by pressing the button on the switch (M), selecting the therapy flow, by increasing or reducing it with
the flow regulator (H), and start treatment preferably seated and in a comfortable position;
• Continue inhalation until the outlet flow is exhausted. If you are not sure that the drug has been finished, you
can suspend the therapy shortly and then restart it. After treatment, there will be a small amount of residual
liquid (approx. 0.5 ml) in the nebuliser chamber.
• The unit has a system that switches it off automatically after 5 minutes from its start. If you need to continue the
therapy, add the drug in the tank again (I) by removing the cover (F). After fitting the cover back on again, start
the unit again by pressing the button on the switch (M).
• WARNING! Never carry out more than two consecutive 5-minute cycles! At the end of a second consecutive cycle,
wait 30 minutes before using again! If you fail to follow these warnings, the unit's safety cut-off may be triggered.
• At the end of the treatment or when necessary, press the button on the switch (M), disconnect the plug of the
detachable power cord (L) from the mains socket, and disconnect the accessory and tube; clean the unit and
accessories according to the instructions in the section "CLEANING AND MAINTENANCE".
CLEANING AND MAINTENANCE
CLEANING AND MAINTENANCE OF THE DRUG TANK
Before cleaning and maintenance of the tank, switch off the unit (if on) by pressing the button on the On/Off
switch (M), unplug the detachable power cord (L) from the mains socket and disconnect the connector of the
detachable cable from the unit.
Warning! Do not wet the unit or put it into water or other liquids.
To clean any residual drug from the tank, use only a soft, clean cloth moistened in warm water. Clean the drug
tank surfaces thoroughly (I) especially in intersection points. Cleaning must be done at the end of every therapy
session. Do not use detergents to clean the device, as they may cause damage.
WARNING! Never clean the drug tank with sharp or rough objects.
WARNING! Never use flammable substances during cleaning and maintenance of the drug tank and unit.
Carefully dry the drug tank and the unit before putting them away.
Before putting the unit away, shut the drug tank with its cover and make sure to close the nebulisation nozzle.
Please clean and check the condition of the drug tank before and after use and do not use the unit if it has clear
signs of damage.
CLEANING AND MAINTENANCE OF ACCESSORIES
Remove all used accessories (mask, mouthpiece, nose fork and tube) and rinse them under running water.
WARNING! Do not put the mask in boiling water. Cold disinfect all parts that come in contact with the patient
using denatured alcohol. Clean and check that the condition of the accessories before and after use. Replace
them if damaged.
Carefully dry all components and the unit before putting it away.
Store the unit in a cool, dry place, away from light and heat.
Never use benzene, thinners or other inflammable chemical substances for cleaning. For increased hygienic safety,
never use the same accessories on more than one patient. Purchase a dedicated kit
for each user (AirProjet kit
WARNING! Never leave the nose fork unattended as it is a small part that, if swallowed by children,
could represent an ingestion and suffocating hazard.
02038406100000).
16