Télécharger Imprimer la page

FLAEM AirPro 3000 Plus Mode D'emploi page 25

Masquer les pouces Voir aussi pour AirPro 3000 Plus:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
- Fill a container of a suitable size to hold all the individual components to be disinfected with a
solution of drinking water and disinfectant, observing the proportions indicated on the packaging
on-
of the disinfectant.
rol
- Completely immerse each individual component in the solution, taking care to avoid the formation
of air bubbles in contact with the components. Leave the components immersed for the period of
time indicated on the disinfectant packaging, and associated with the concentration chosen for the
preparation of the solution.
- Recover the disinfected components and rinse them thoroughly with lukewarm drinking water.
- Dispose of the solution according to the disinfectant manufacturer's instructions.
method B: Disinfect accessories by boiling them in water for 10 minutes; use demineralised or distilled
water to avoid hard water deposits.
method C: Disinfect the accessories in a hot bottle steamer (not microwave). Carry out the process
y.
strictly according to the hot bottle steamer's instructions. For disinfection to be effective, choose a
steriliser with an operating cycle of at least 6 minutes.
If you also want to carry out sterilisation, skip to the Sterilisation section.
After disinfecting the accessories, shake them vigorously and place them on a paper towel, or
alternatively dry them with a hot air jet (e.g. hair dryer).
At the end of each use, store the device complete with its accessories in a dry and dust-free place.
Sterilisation
Device: Steam steriliser with fractionated vacuum and overpressure compliant with EN 13060.
Execution: Package each individual component to be treated in sterile barrier system or packaging
in accordance with EN 11607. Place the packaged components in the steam steriliser. Perform
the sterilisation cycle in accordance with the instructions for use of the equipment by selecting a
temperature of 134°C and a time of 4 minutes first.
Storage: Store sterilised components according to the instructions for use of the system or sterile
barrier packaging chosen.
The sterilisation procedure was validated in accordance with ISO 17665-1.
At the end of each use, store the device complete with its accessories in a dry and dust-free place.
ns.
nd).
the
B2.
ods
g
en
or
the
ure
um
25

Publicité

loading

Ce manuel est également adapté pour:

P0712em f3000Rf7-2