Guidance and manufacture's declaration
all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic
immunity
The SPO 55 Pulse Oximeter is intended for use in the electromagnetic environment
specified specified below. The user of SPO 55 Pulse Oximeter should assure that it is
used in such an environment.
Immunity
IEC60601 test
test
level
Electrostatic
± 6 KV contact
discharge
± 8 KV air
(ESD) IEC
61000-4-2
–
electromagnetic immunity for
Compliance
Electromagnetic envi-
level
ronment-guidance
± 6 KV contact
Floors should be wood,
± 8 KV air
concrete or ceramic tile. If
the floor is covered with
synthetic material, the
relative humidity should
be at least 30%.
GB/IE/NI
45