17
Technical Information
Data Exchange Protocols
Error Code Table
Tab. 4
CODE
00E000
00E001
00E002
00E003
00E004
00E005
00E006
00E009
00E010
RS-232 Protocol 2 Communication Descriptions
Transmission Channel Characteristics
Tab. 5
levels:
baud rate:
character length:
parity:
stop bit:
protocol:
190/209
DESCRIPTION
No error
Output current > 15 A
Pump not connected
Output current > 1.5 A for more than 15 sec.
Bearing temperature > 60 °C
Heat exchanger temperature > 60 °C
Normal not reached within run up time
Vp voltage 8V lower than nominal for more than 5 sec.
Soft start ramp not ended within the expected time
RS-232
9600/4800 programmable by W7 jumper on the board
8 bits
none
1 bit
master (PC) / slave (converter)