WARNING
Do not exceed 60 seconds of contact
between this probe and any circuit with
voltages ≥8 kV DC, ≥5 kV
kV
pk-pk
such measurements.
3
Proper compensation of the probe is
required to assure amplitude accuracy of
the waveform being measured by matching
the probe to the oscilloscope's input
capacitance. Compensation should be
adjusted whenever the probe is connected
to, or transferred between, oscilloscopes.
The procedure is the same as for a x10
passive probe.
187
AC. Wait for 5 minutes between
Frequency Compensation
Connect the probe to the
oscilloscope.
Connect probe tip to square-wave
generator.
Adjust the square wave generator for
10 V amplitude.
Adjust the oscilloscope for a time
base of 500 µs/div.
Turn trimmer (Figure 1) to adjust the
compensation on the probe making a
flat pulse top.