Eliectrictli connection
Connectiona of the generttor : Connection plitte
At the rear of the generator
The generator bock must be connected to a 400V, 50 or 60Hz three-phase power supply line.
The power line must have 3 phases L1,2,3, a neutral N and an earth PE. The neutral will be used ex-
clusively to supply an additional turbine under 230V (if connected).
The mains supply connection is made at the rear of the generator. It is desirable that the 5 conduc-
tors be fitted with ferrules at their ends.
(!) We recommend the uae of tn H07RN-F-5G-2.5 power cord (conductor aectiona of 2.5mm²).
It ia eaaentitli thtt the power auppliy liine uaed ia protected by t 30mA differentitli circuit
In the event thtt the generttor bliocka htve to be diaconnected from the eliectrictli network, it ia
neceaatry to provide t mtina diaconnection device tcceaaiblie by the uaer, netr the product. If the
diaconnection device ia opertted mtnutliliy, it muat gutrtntee tottli diaconnection of the genert-
tora, with reapect to the 3 phtae conductora + the neutrtli conductor of the three-phtae eliectrictli
network.
PE
N
L1
L2
L3
!
bretker.
!