Troubleshooting - Vermeiren Albatros Mode D'emploi

Masquer les pouces Voir aussi pour Albatros:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
4.6

Troubleshooting

Problem
Noisy sound of the moving
parts (example: cantilever
arm).
Actuator makes an unusual
sound.
Patient hoist doesn't lift
anymore.
Lowering the patient in
case of emergency doesn't
work.
Battery cannot be charged
C ause of problem
Lubrication is necessary.
The actuator is damaged.
1. Electric actuator is
damaged.
2. The user weight is too
high.
3. Manual control or
actuator aren't
connected.
4. Battery not connected
or not connected
properly to control box.
5. Battery too low.
6. Emergency stop is
activated.
There is required a minimum
weight load to lower the
patient.
1. Batteries not fitted
correctly to control box.
2. Mains cable defect.
3. Batteries are defected.
4. Control box with
included charger is
defected.
Page 21 
Problem solving
Lubricate the moving parts.
(Do not lubricate the actuator!)
Replace the actuator or contact your
specialist dealer.
1. Replace the actuator.
2. Use a patient hoist with higher
maximum user weight.
3. Check connections. If
necessary connect manual
control or actuator.
4. Check battery is installed or the
connection of the battery.
5. Charge the batteries or replace.
6. Deactivate the emergency stop.
If the patient weight is too low pull
down slightly on the cantilever arms
or adjust the lowering speed.
1. Check batteries are
connected to the control box.
2. Change mains cable.
3. Replace battery.
4. Replace control box with
included charger.
EN
Albatros
2010-12

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Albatros 1

Table des Matières