FORMat Subsystem
FORMat commands specify the format for transferring measurement data.
FORMat[:DATA] ASCII|REAL
FORMat[:DATA]?
Specifies the format of the returned data. This is used by queries that can return a block of data. ASCII
returns data as ASCII bytes in numeric format as appropriate. The numbers are separated by commas.
REAL returns data in a definite length block as IEEE single precision floating point values. In this case
the 4 bytes of each value can be returned in either big-endian or little-endian byte order, determined
by the FORMat:BORDer setting.
Parameter
ASCII|REAL, *RST ASCII
Sets the data format to ASCII: FORMat ASCII
The data format is used by a small sub set of queries that can return large quantities of data.
l
FORMat:BORDer NORMal|SWAPped
FORMat:BORDer?
Specifies how binary data is transferred. This only applies when the FORMat:DATA is set to REAL.
NORMal transfers data in normal order. The most significant byte is returned first, and the least
significant byte is returned last (big-endian). SWAPped transfers data in swapped-byte order. The
least significant byte is returned first, and the most significant byte is returned last (little-endian).
Parameter
NORMal|SWAPped, *RST NORMal
Sets the data transfer to Swapped: FORM:BORD SWAP
The byte order is used when fetching real data from SCPI measurements.
l
Keysight RP7900 Series Operating and Service Guide
Typical Return
ASCII or REAL
Typical Return
NORM or SWAP
5 SCPI Programming Reference
205