Clear Output Protection
If an over-voltage, over-current, over-temperature, power-fail condition, power-limit condition,
protection condition, or inhibit signal occurs, the output is disabled. The appropriate operating status
indicator on the front panel will be on. To clear the protection function and restore normal operation,
first remove that condition that caused the protection fault. Then, clear the protection function as
follows:
Front Panel Menu Reference
Select Protect\Clear
Select Clear.
The output is re-enabled once the output protection function is cleared.
Keysight RP7900 Series Operating and Service Guide
4 Using the Regenerative Power System
SCPI Command
To clear a protection fault:
OUTP:PROT:CLE