Undesired effects of
overlapping light fields
Optical safety
Incorrect use
Operation and
maintenance manual
WARNING – Possibility of tissue dehydration and damage
In the event of overlapping lamps, a temperature increase would ensue
in the patient area with consequent risk of dehydration and tissue
damage.
In case of a reduction in blood flow with start of tissue dehydration,
reduce light intensity.
2.2 Safety conditions (secondary effects)
CAUTION – Possibility of glare
- Do not direct the light source into the patient's and/or operator's eyes.
- When Product use is restricted to the face (maxilla-facial surgery,
plastic surgery, ear-nose-throat surgery) the patient's eyes must be
covered with adequate protection.
Failure to follow such precautions could cause glare and potential
damage to the retina.
CAUTION – Do not place objects on Product
Never place and/or hang anything on the Product.
Failure to follow such precaution could result in such objects falling in
the operating area.
CAUTION – Possibility of damaging Product
- Never hang on the Product with the body weight of a person.
Failure to follow such precaution could damage the Product structure.
-Never cover the head of the Product during operation to prevent
overheating.
- Avoid the Product parts colliding with one another or other nearby
equipment.
Knocks could cause the detachment of plastic parts or paint from the
Product which could fall in the patient area
MU_024
04/02/16
Rev.2
Page 6 of 28