Power
frequency
(50Hz)
magnetic
field
IEC 61000-4-8
Guidance and manufacture's declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration-electromagnetic immunity
The CMS50Q1 Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of CMS50Q1 Pulse
Oximeter should assure that it is used in such an environment.
Immunity
IEC60601
test
test level
Radiated
3V/m
RF
80MHz
ICE
to
61000-4-
2.5GHz
3
3A/m
3A/m
Compliance
level
3V/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment
Electromagnetic
environment -guidance
Portable and mobile RF
communication equipment
should be used no closer to any
part of the CMS50Q1 Pulse
Oximeter, including cables,
than the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter.