Télécharger Imprimer la page

LaMotte TRACER Mode D'emploi page 2

Masquer les pouces Voir aussi pour TRACER:

Publicité

Les langues disponibles

Les langues disponibles

„ QUICK CALIBRATION PROCEDURE
1. Fill sample cup to 20 mL line with included 3000 ppm Salt Calibration
Standard (6005).
2. Remove cap.
3. Press ON/OFF button to turn meter on.
4. Put electrode in standard. Stir to remove air bubbles.
5. Be sure meter is in Salinity mode before starting calibration. See Changing
the Measurement Function on page 5.
6. Press and hold CAL button for 2 seconds until "CAL" appears and display
flashes.
7. "SA" and "End" will appear briefly, then release CAL button.
8. Rinse and dry the electrode. The meter is ready to use.
Tips: Repeat Quick Calibration monthly or more frequently if needed. For more
details see Calibration on page 7.
CONTENTS
SAL/EC/TDS TRACER PockeTester Kit, Code 1749/1749-KIT-01; Sample Cup with
cap, Code 1745-1; Salinity Calibration Standard, 3,000 ppm, 30 mL (1749) Code
6005-G; Salinity Calibration Standard, 3,000 ppm, 120 mL (1749-KIT-01) Code
6005-J
PARTS & ACCESSORIES
EC/TDS/SAL Replacement Electrode
Weighted Stand w/Sample Cups (5)
Sample Cups w/caps (24)
Conductivity Standard, 84 µS, 30 mL
Conductivity Standard, 1413 µS, 30 mL
Salinity Standard, 3,000 ppm, 30 mL, 120 mL
3' Cable for on-site testing
*Reagent is a potential health hazard.
READ SDS www.lamotte.com.
Emergency Information: Chem-Tel USA 1-800-255-3924
International, call collect, 813-248-0585.
SPECIFICATIONS
Display
Conductivity Ranges
TDS Ranges
Salinity Ranges
2000 count LCD with Bar Graph
0 to 199.9 µS, 200 to 1999 µS, 2.00 to 19.99 mS
0 to 9,999 ppm, 100 to 999 ppm or mg/L, 1,000 to 9,999
ppm
0 to 9,999 ppm, 100 to 999 ppm, 1,000 to 9,999 ppm