Specifications
Connection.............................................................
Power supply..........................................................
Data transmission parameters...............................
Max. cable length...................................................
Number of transmissions per second....................
Data transmission format.......................................
Data format
Data
Sign : « + », « - », ou « space »
E1-En: integral number
F1-Fn: decimals
n:
depends on used unit and resolution
Errors
0: sensor error (e.g. speed, scale distance)
2: parity error (duplex instruments only)
Identification
XX:
Manufacturer
Instr:
203, 235, 233, etc
OPT1: version
OPT2: additional options (according to instrument used)
Note: The transmission of the identification is done only after having switched ON
the instrument (simplex instruments)
Remote commands
Format
C1-Cn:
command with 2 or 3 characters
S1-Sn: 0/1 :
command disabled/enabled
? :
status request
+/- XXX.YYY: entering of numerical values
connector 2x5 poles, pitch 2.54mm
pin 9 of the peripheral unit, +5 - +9V DC
Digimatic compatible
3m
4-8/s (depends on the instrument connected)
Digimatic compatible
[ Sign ¦ E1-En ¦ «.» ¦ F1-Fn ¦ CR ]
[ «ERR» ¦ Number ¦ CR ]
1: incorrect command
3: exceeding of the measurement range
[ «XX» ¦ Instr. ¦ «.» ¦ OPT1 ¦ {«.» ¦ OPT2} ¦ CR ]
[¦ C1-Cn ¦ { S1-Sn } ¦ CR ]