Télécharger Imprimer la page

IDEAL 61-605 Manuel D'instructions page 7

Masquer les pouces Voir aussi pour 61-605:

Publicité

Les langues disponibles

Les langues disponibles

Ranges & Accuracies:
Function
AC Voltage 200.0mV/2.000V/20.00V/200.0V/600V 4.0% + 5 digits 600 VDC or AC rms
(50/60 Hz)
DC Voltage 200.0mV/2.000V/20.00V/200.0V/600V 2.0% + 2 digits 600 VDC or AC rms
200.0Ω
2.000kΩ/20.00kΩ/200.0kΩ
Resistance
2.000MΩ
20.00MΩ
Audible indication < 25Ω
Continuity
Response time: 500ms
Test current: 1.2mA
Diode
Open circuit volt: 3.0 VDC typical
Input Impedance is 9.1MΩ (20-600V AC/DC), 10MΩ (2V AC/DC), 100MΩ
(0-200mV AC/DC)
Double Insulation
Instrument has been evaluated and complies with insulation category II (overvolt-
age category II). Pollution degree 2 in accordance with IEC-644. Indoor use.
Range and
Resolution
~
7
Accuracy
2.0% + 5 digits
2.0% + 4 digits
450 VDC or AC rms
3.0% + 4 digits
5.0% + 5 digits
N/A
450 VDC or AC rms
3.0% + 3 digits
Overload
Protection

Publicité

loading