Measuring Continuity and Resistance
CONTINUITY
Set the selector switch to the position
Check the circuit to be measured with the contact tips. Audible continuity beeping will
be active for a resistance ≤ threshold selected by programming (40 Ω by default).
RESISTANCE
Starting from the Continuity position, press the yellow button (secondary functions)
to select the Resistance function.
Check the circuit to be measured with the contact tips.
Range switching is automatic.
Calibrated range overloading is indicated on the display by
Possibility of compensating for the resistance of the power cables by a long press
on the
button and a shorting of the cables.
HOLD
Display Range
Measurement Range
Precision
Resolution
Voltage in Open Circuit
Protection
400 Ω
0,0 - 399,9 Ω
1,5 % ± 2 points
1.5 % ± 2 points
0,1 Ω
3,2 V
500 V AC or 750 V (DC or Peak)
(Ohmmeter)
.
OL
4,000 Ω
400 - 3,999 Ω
1 Ω
3,2 V
17