Safety
Never take measurements of resistances and diodes on a powered
circuit. Never use on voltage networks higher than 600 Vrms.
Before taking any measurement, make sure the power cables and the selector switch
are positioned correctly.
Never open the tester case before all the electrical sources have been disconnected.
Never connect to the circuit to be measured if the tester case is not closed correctly.
Use accessories complying with standards of safety (EN 61010-2-031), 600 V
minimum voltage and overvoltage category III.
Specifications
Overall dimensions
Weight
Opening of alligator clips
1 battery
Autonomy
NB : Restarted by passing over the selector switch's OFF position or by pressing any button.
NB :
NB :
NB :
NB :
Battery discharged indication: A flashing battery symbol indicates an
autonomy of less than 1 hour.
Operating temperature
Storage temperature
Relative humidity in operation :
Relative humidity in storage
Conformity to standards
Degree of protection
Self extinguishing
Radiated fields
Free fall
Vibrations
Shocks
Electrical shocks
Electrostatic discharge
Rapid transients
Automatic or manual measurement mode selection (AC or DC) (manual mode selected
by pressing the yellow button)
Automatic measurement range selection
Automatic tester turn-off system
Recording functions of the MIN - MAX - PEAK values
Display Hold by pressing the HOLD button (blue button)
Liquid crystal display, 4,000 measuring points (limited to 600 V), 8 mm digit height
Display refresh rate : 400 ms
Backlighting controlled by a manual button (green button)
Automatic display of polarity (+ and -) for DC measurements and PEAK values
Display overLoad indication : OL
DC zeroing to correct for variations
Automatic compensation for resistance of measuring cables in ohmmeter function (W zero)
Measuring of currents by clamping
V-Live function. «BEEP» if voltage > 30 V
:193 mm x 70 mm x 37 mm
:260 grams
:26 mm
:9 volts (type 6LF22)
:100 hours or 6,000 measurements of
1 minute each. Automatically turns off
after 10 minutes without the selector
switch or a function button being activated.
A fixed battery symbol implicates the battery
has to be changed.
:
0° C to + 50° C
:
- 20° C to + 70° C
< 80 % RH
:
< 90 % RH
(up to + 45° C)
: EN 61010.1 (Ed. 95) Class II (double insulation)
Degree of pollution 2
Installation category III
Service voltage 600 V
: EN 6110.2.032 (Ed. 93)
: EN 61326 (Ed. 98)
according to EN 60529(Ed.92)
:
IP 40
K 04
according to EN 50102 (Ed. 95)
:
UL 94 V0
:
3 V/m
(according to IEC1000.4.3)
:
1 m
(according to IEC 68.2.32)
:
.75 mm
(according to IEC 68.2.6)
:
.5 Joule
(according to IEC 68.2.27)
:
6 kV
(according to IEC 1000.4.5)
:
4 kV class 2
(according to IEC 1000.4.2)
:
2 kV
(according to IEC 1000.4.4)
EN