10
Rotors for the Microcentrifuge 24
Only use rotors that are intended for use with the corresponding centrifuge.
Please note the manufacturer's information on the centrifugation resistance of the sample tubes used (maximum g-force).
10.1
Rotor Standard 24 AT
Aerosol-tight fixed-angle rotor for 24 tubes
Max. g-force:
Max. rotational speed:
Rotor Standard 24 AT Max. load (adapter, tube and contents):
Tube
Tube
Capacity
Tubes per adapter/rotor
PCR tube
0.2 ml
1/24
Micro test tube
0.4 ml
1/24
Micro test tube
0.5 ml
1/24
Microtainers
0.6 ml
1/24
Micro test tube
1.5 ml/2 ml
–/24
21300 × g
15060 rpm
24 × 3.75 g
Adapter
Bottom shape
Order no.
Diameter
(international)
Conical
Ø 6 mm
8030-0014
Conical
Ø 6 mm
8030-0015
–
Ø 8 mm
8030-0013
–
Ø 8 mm
8030-0013
–
Conical
Ø 11 mm
Rotors for the Microcentrifuge 24
Max. g-force
Max. rotational speed
Radius
21300 × g
15060 rpm
6.3 cm
21300 × g
15060 rpm
8.4 cm
21300 × g
15060 rpm
7.3 cm
21300 × g
15060 rpm
8.4 cm
21300 × g
15060 rpm
8.4 cm
Microcentrifuge 24 · Manual ·
Microcentrifuge 24
31