Operation
Microcentrifuge 24
5.6.2

End of centrifugation

Press the STOP/OPEN key to end centrifugation before the set time.
• After completion of the set time, the centrifuge stops automatically.
• After the run, the lid opens automatically.
• During the braking process, the elapsed running time flashes on the display.
• If the speaker is switched on, a signal sounds when the rotor has stopped.
5.6.3

Centrifuging in continuous operation

Setting continuous run
1. In order to centrifuge without any time limits, use the TIME arrow keys to select the setting ∞ ( below 10 s or  above 9:59 h).
2. Set the rotational speed (rpm) or g-force (rcf) with the SPEED arrow keys.
3. To start the centrifugation run, press the START key.
• The
symbol is flashing when the rotor is running.
• The cycle time is counted up.
• Current g-force (rcf) and/or rotational speed (rpm).
5.6.4

Short run centrifugation

Short run centrifugation takes places up to the maximum rotational speed of the rotor used (15060 rpm).
 Keep the SHORT key pressed.
The centrifuge accelerates up to the maximum rotational speed of the rotor used and continues to run at maximum speed.
 Release the SHORT key.
The centrifuge stops.
5.6.5

Setting the alarm volume

Prerequisites
The lid is open.
 Touch and hold the START key and set the desired volume using the TIME arrow keys.
20
· Microcentrifuge 24 · Manual