NOTE:
10 short beeps will sound when the temperature is higher than 37.5°C (99.5°F)
in order to alert the patient that he/she may have fever. In order to assure the
accurate reading, please wait at least 30 seconds after 3-5 continuous
measurements.
Accumulation of ear wax on the probe tip can result in less accurate temperature
readings or cross infection between users. Therefore, it is essential that a clean
probe tip is used before each measurement. For cleaning, please follow the
instructions in the (Cleaning and Storage) section.
After cleaning the probe tip with alcohol, it is necessary to wait 5 minutes before
taking the next measurement, in order to allow the thermometer to reach its
operating reference temperature.
• For an infant, it is best to have the child laying flat with his head
sideways so the ear is facing upwards. For an older child or adult,
it is best to stand behind and slightely to the side of the patient.
• Always take
the temperature in the same ear, since the temperature
readings may be different from the right ear and left ear.
• Please wait for a few
minutes to take the ear temperature after
sleeping.
• In the following situations it is recommended that three temperatures
in the same ear be taken and the highest one taken as the reading
1-New born infants in the first 100 days.
2-Children under three years of age with a compromised immune
system and for whom the presence or absence of fever is critical.
3-When the user is learning how to use the IR thermometer for the first
time until he/she has familiarized himself/herself with the instrument
and obtains consistent readings.

7.Changing from Fahrenheit to Celsius and Vice-versa

The Digital Infrared Ear Thermometer can display temperature measurements
In either Fahrehelt or Celsius. To switch the display between °C and °F, simply
turn OFF the unit, press and hold the Start button for 5 seconds; When you stop
pressing the Start button after 5 seconds, the current measurement scalef
C or °F icon) will be flashed on the display. Toggle the measurement scale
between °C and °F by pressing the start button again. When the measurement
scale has been chosen, wait for 5 seconds and the unit will enter the ready for
measuring mode automatically.