Optima Auto Link / English
9.3. Others
Q1 Power is not ON
‧ Check if the plug is connected to mains.
‧ Check for a blown fuse.
Q2 Low Pressure Alarm is on
‧ Check if the Quick Connector is tightly secured.
‧ Check if all tubing connections along with mattress are secured.
‧ Check if the air hoses are kinked or broken.
‧ Check if the CPR knob is tightly closed.
Q3 Power Failure Alarm is on
‧ Check if the power is suddenly shut down.
‧ Check if the power cord is connected properly.
Q4 Patient is bottoming out
Pressure setting might be inadequate for the patient, increase comfort range 1 level higher and wait for a
few minutes for best comfort.
Q5 Mattress form is loose
‧ Check if all the snap buttons or straps of mattress are all securely fastened.
‧ Check if the mattress is fixed to the bed frame by straps.
Q6 No air produced from some air outlets of the air tube connector
This is normal since there is alternating mode. Air outlets take turns to produce air during their cycle time.
Q7 System malfunction (Code 11 or 12)
Check if there is only one or two knobs at the foot-end of the mattress remained open.
If the above information does not solve your problems, please contact your local agent directly.
They might require a technician to take care of the problem.