BYK-Gardner GmbH
7 Maintenance
Disconnect power cord before cleaning the light booth or replacing lamps. Lamps are hot
after use! Let them cool down first.
Cleaning
To clean the neutral gray surface or diffuser, use a soft, damp cloth with a water based
mild soap or detergent. Allow to air dry. Release the latch to reach the inner side of the
diffuser.
Lamp replacement procedure
Remove the front panel of the light booth. Release diffuser to reach lamps. To protect the
lamps, only touch them with gloves.
Lamp type
Daylight (D65)
Horizon (HZ) /
Incandescent (A)
CWF / TL84 / U30 /
Ultra-violet (UV)
301 300 039 - 2303
Lifetime
Replacement note
600 hrs
Please contact our technical service for re-
placement.
600 hrs
Please push one end of the lamp into the
spring-loaded socket and release the other
end.
After replacement perform Self-Calibration
as described on the next page.
2000 hrs
Please rotate the tube 90 degrees to loosen it.
byko-spectra pro
11