MB35
3.3 Temperature Calibration
Note: Temperature Calibration Kit is required.
TEMP CAL
INSERT CAL FIXTURE
CLOSE COVER
ABORT CALIBRATION
TEMP CAL
CALIBRATION READY
PRESS TARE/ACCEPT
ABORT CALIBRATION
Adjust to thermometer reading
/
TEMP CAL 100° C
CURRENT TEMP:
100° C
TM TO CAL POINT:
15:00 MIN
ADJ CAL READING:
104°
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 160° C
CURRENT TEMP:
160° C
TM TO CAL POINT:
0:00 MIN
ADJ CAL READING:
163°
ACCEPT NEW CAL
ABORT CALIBRATION
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
All manuals and user guides at all-guides.com
ANALYZER SETUP
LOCK OUT
WEIGHT CAL
TEMP CAL
/
LANGUAGE
BEEPER
TIME-DATE
TEMP CAL 100° C
CURRENT TEMP:
TM TO CAL POINT:
ADJ CAL READING:
TEMP CAL 100° C
CURRENT TEMP:
TM TO CAL POINT:
C
ADJ CAL READING:
ACCEPT NEW CAL
ABORT CALIBRATION
15
TEMP CAL 160° C
CURRENT TEMP:
TM TO CAL POINT:
C
ADJ CAL READING:
ACCEPT NEW CAL
ABORT CALIBRATION
Open the cover, allow unit to cool, remove the calibration fixture,
replace the pan support and pan.
REMOVE PAN SUPPORT
ABORT CALIBRATION
¾ C
15
15:00 MIN
100°
C
ACCEPT NEW CAL
ABORT CALIBRATION
100° C
15:00 MIN
104° C
Adjust to thermometer reading
TEMP CAL 160° C
CURRENT TEMP:
160° C
TM TO CAL POINT:
0:00 MIN
ADJ CAL READING:
160°
C
EN-9
TEMP CAL
TEMP CAL 100° C
CURRENT TEMP:
100° C
TM TO CAL POINT:
15:00 MIN
ADJ CAL READING:
100°
C
ACCEPT NEW CAL
ABORT CALIBRATION
/
160° C
0:00 MIN
163°
C
ACCEPT NEW CAL
ABORT CALIBRATION