The applied colors:
10 μl pipettes – red,
50 μl, 200 μl pipettes – yellow
300 μl pipettes - green
1.2. WORK SAFETY INSTRUCTIONS
Used symbols:
WARNING! Risk of injury
CAUTION: Potential damage to the device or pipetting
errors
Long-lasting and faultless working of the pipette
depends to a large extent on its operation. Read the
pipette operating instructions carefully and comply with
the principles included.
CAUTION:
• Pipette is designed to work with the tips only. Do not
aspi rate liquids without tips attached. Aspirated liquid
should not get into the pipette because this may dam-
age it.
• Use only the pipette with the tips attached.
• One time use of a tip guarantees safety and eliminates
possible contamination of the aspirated liquid.
• Keep the pipette clean; do not cleanse it with chemically
aggressive substances (e.g. acetone).
• Do not hold the pipette horizontally if there is liquid in
the tip.
• Only use pipettes in accordance with the manufacturer
instruction to ensure the proper parameters of pipettes.
• In case of improper operation of the pipette, clean the
pipette carefully following the instructions given or
send it to the service representative.
WARNING!
• Follow the laboratory work safety regulations.
• Exercise extreme caution while pipetting chemically
aggressive substances. Use the protective equipment,
glasses and gloves.
• Never direct the pipette with the tips attached towards
yourself or anyone else when there is liquid in it.
• Use only the spare parts and accessories recommend-
ed by the manufacturer.
3
MULTICHANNEL
1.3. TECHNICAL PARAMETERS
One Touch Pro pipettes are high quality laboratory
instruments, which provide the highest measurement
accuracy and precision.
Accuracy errors and repeatability of liquid volume mea-
surements depend on quality of the applied tips. The
errors provided, have been determined with the use of
SORENSON tips by means of gravimetric method based
on minimum 10 samples of distilled water with tempera-
ture of 20°±1°C under the EN ISO 8655 standard.
Volume
Accuracy
Model
Cat.no.
[μl]
± 8,0
Min
1
OTP8-10
177060
± 4,0
5
OTP12-10
177100
± 2,0
Max
10
± 4,0
Min
5
OTP8-50
177070
± 3,0
25
OTP12-50
177110
± 1,6
Max
50
± 3,0
Min
20
OTP8-200
177080
± 1,5
100
OTP12-200
177120
± 1,0
Max
200
± 1,6
Min
50
OTP8-300
177090
± 1,2
150
OTP12-300
177130
± 1,0
Max
300
Table 1
Only application of SORENSON tips guarantees accura-
cy of measurements at the level shown in the table
above.
The pipette design enables the user to perform calibration
process according to the rules presented in section 5.
2. OPERATION
2.1. SETTING THE VOLUME
Setting the volume is performed by the volume adjust-
ment knob (fig. 1B).
ENGLISH
Precision
Increment
Fits to tips
[%]
[%]
[μl]
[μl]
± 6,0
± 2,0
0,01
10
± 1,2
± 2,5
± 1,2
0,05
± 0,6
200
± 1,5
± 0,8
0,2
± 0,6
± 1,5
± 1,0
0,5
300
± 0,6
4