Calibration/adjustment with internal weight: Model TP-303I
Menu setting 1 5 2 required.
Step
1. Calibrate/adjust
the unloaded balance
2. Tare the unloaded balance
3. Start calibration/adjustment
4. Move the selector for the
motorized weight into the
"calibration/adjustment"
position
5. Calibration/adjustment
is performed
("CC" flashes on the display)
6. Remove weight:
Return the selector to the
initial position
7. The internal weight is removed
from the weighing system
Key (or instruction)
[TARE]
[CAL]
Display/Printout
0.0
g
c
c
cC
cc
0.0
g
11