Calibration/Adjustment
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance,
– the balance is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed.
Otherwise, the weight required for calibration/adjustment is displayed.
You can use any of the following weight units to calibrate/adjust:
g, kg*, lb (menu code 1. 4. x)
You can block calibration/adjustment of the balance:
– Select menu code 1. 5. 3
External calibration/adjustment
Step
1. Tare the balance
2. Begin calibration
Calibration weight is displayed
without weight unit (here: 1000 g)
3. Place the indicated calibration
weight on the balance
After calibration, the calibration
weight is displayed with wt. unit
4. Remove the calibration weight
* = not on models with a readability of 0.1 mg
10
Key (or instruction)
[TARE]
[CAL]
Display/Printout
0.0
g
+
1000.0
1000.0
+
1000.0
g
0.0
g